System |
|
Scanner
Description |
Pencil
beam DXA (dual energy x-ray absorptiometry) |
|
Drive
System |
Stepper
motors, with Kevlar reinforced drive belts |
|
X-ray
Source |
Stationary
anode x-ray tube, air-cooling; 100kV constant potential,
1.3 mA anode current (1.0 mA Japan), focal spot size:
0.5 mm |
|
Filtration |
8-level
automated samarium filtering (K-edge = 46.8 keV) -
minimum filtration is 2.7mm aluminum equivalent |
|
X-ray
Detectors |
2
NaI scintillation detectors in pulse counting mode |
|
Power |
100-120,
220-240 V~,50/60 Hz, 700 VA maximum |
|
Operating
Temperature |
60°-104°(F)
(15°-40°(C)) |
|
Relative
Humidity |
Up
to 80% non-condensing |
|
Footprint |
|
Dimensions |
Recommended
dedicated floorspace: 84"L x 84"W
Tabletop size: 72"L (1828 mm) x 35.5"W (902
mm) at 26"H (660 mm)
Overall size: 72"L (1828 mm) x 48"W (1220 mm)
x 49"H (1245 mm)
Table surface to arm (maximum patient thickness):
22"(559 mm) |
|
Weight |
400
lbs. (181 kg) |
|
Performance |
|
Scan
Sites |
AP
Spine, Hip, Forearm and Lateral Spine: optional scan
modalities: Research*, Small subject*, Body composition |
|
Scan
Time |
<1.5
minutes AP Spine, <2.0 minutes Hip, <3.0 minutes
Forearm, <4.0 minutes Lateral Spine |
|
BMD
Precision |
1.0%
- AP Spine, 1.2% - Hip, 0.8% - Forearm, 2.4% - Lateral
(2 vertebra) |
|
Accuracy |
1%
(based on hydroxyapatite phantom) |
|
Spatial
Resolution |
|
|
AP
Spine |
Line
Spacing - 1.5 mm (1.0 mm selectable)
Point resolution - 1.5 mm (1.0 mm selectable) |
|
Proximal
Femur |
Line
Spacing - 1.0 mm
Point resolution - 1.0 mm |
|
Forearm |
Line
Spacing - 1.0 mm
Point resolution - 1.0 mm |
|
Lateral |
Line
Spacing - 1.0 mm
Point resolution - 1.0 mm |
|
Scan
Dose |
<1.0
mRem typical (AP Spine, Hip, Forearm) and <2 mRem
typical (Lateral) |
|
Operation |
|
Active
Scan Area |
25.5"x
30" (650 mm x 762 mm) - with off-axis scanning
capability |
|
Patient
Marking |
Laser
diode, red, class II (less than 1 mW power) |
|
Patient
Positioning Aids |
AP
Spine positioning block, Hip sling with foot separator,
Lateral and Forearm positioning aids, scout scan, and
laser marking system |
|
Calibration |
Automatic
with supplied 77-Step calibration standard & QC
phantom |
|
Computer
Workstation |
|
Computer |
IBM
compatible PC computer, with 366MHz InTel® CPU (366MHz
laptop optional), 32 MByte RAM, 1.44 MByte floppy disk
drive, Pointer device (mouse), CD-ROM, fax/modem |
|
Operating
System |
DOS,
Windows® 98 |
|
Hard
Disk |
6.4
GByte minimum, internal (4.0 GByte minimum with laptop
option) |
|
Display |
15"
Super VGA color (LCD screen only with laptop option).
Optional flat screen monitor or 17" SVGA monitor
available |
|
Printer |
HP
Color Desk Jet |
|
Tape
Backup |
HP
14 GByte Backup Tape Drive with tape (not available with
laptop option) |